HJ734-2014 Analysis of volatile Organic Compounds from stationary pollution sources (Thermal desorption-GCMS)

Time:2023.02.20
Publisher:Admin
Number:2654
2023.02.20
Admin
2654

Abstract

       Referring to HJ 734-2014 Analysis of Volatile Organic Compounds from Stationary Sources, 24 kinds of volatile organic compounds were detected by GC-MS. Volatile organic compounds were qualitatively and quantitatively by internal standard method according to retention time, mass spectrometry and characteristic ions.

1. Configuration scheme

10230531162027.jpg

2 Test the conditions

Injection method: automatic thermal desorption instrument

Injection mode:Split;Split ratio: 10:1

Column: RTX-624, 60m×0.25mm×1.4μm constant-current mode column flow: 1ml/min

Column oven heating procedure: 30℃ (5min) 6℃/min 140℃ (0min) 15℃/min 220℃ (3min)

GC Inlet: 200℃; Ion source: 230℃; Transmission line: 250℃;

Solvent delay: 8min;

SCAN mode: Full scan (SCAN) or select ion (SIM); Scanning range: 35~270amu.

3、Test results

3.1 Qualitative results of volatile organic compounds

2rtyu.png

Figure 1 Standard spectrum of 100mg/L volatile organic compounds

30230531162149.jpg

3.2 TIC diagram of standard samples of different concentrations

(Contact us for more details)

  • Adress
    401, Zone 16A, Jintong International Industrial Park, No. 8 Xihu Road, Wujin High tech Zone, Changzhou City, Jiangsu Province

  • Tel
    0519-86223629

  • Email
    PANYU@pannatek.com

  • WeChat
    Scan Code

+
MESSAGE
You can fill out the information below and leave us a message, and we will reply to you as soon as possible.
Corporate Name*
Product*
Name*
Tel*
MESSAGE
Welcome to visit
Changzhou Panyu Instrument Co., Ltd. Official Website
For more information, please consult:0519-86223629